Drug susceptibility testing of clinical isolates of streptococci and enterococci by the Phoenix automated microbiology system
BMC Microbiology volume 7, Article number: 46 (2007)
Abstract
Background
Drug resistance is an emerging problem among streptococcal and enterococcal species. Automated diagnostic systems for species identification and antimicrobial susceptibility testing (AST) have become recently available. We evaluated drug susceptibility of clinical isolates of streptococci and enterococci using the recent Phoenix system (BD, Sparks, MD). Diagnostic tools included the new SMIC/ID-2 panel for streptococci, and the PMIC/ID-14 for enterococci. Two-hundred and fifty isolates have been investigated: β-hemolytic streptococci (n = 65), Streptococcus pneumoniae (n = 50), viridans group streptococci (n = 32), Enterococcus faecium (n = 40), Enterococcus faecalis (n = 43), other catalase-negative cocci (n = 20). When needed, species ID was determined using molecular methods. Test bacterial strains were chosen among those carrying clinically-relevant resistance determinants (penicillin, macrolides, fluoroquinolones, glycopeptides). AST results of the Phoenix system were compared to minimal inhibitory concentration (MIC) values measured by the Etest method (AB Biodisk, Solna, Sweden).
Results
Streptococci: essential agreement (EA) and categorical agreement (CA) were 91.9% and 98.8%, respectively. Major (ME) and minor errors (mE) accounted for 0.1% and 1.1% of isolates, respectively. No very major errors (VME) were produced. Enterococci: EA was 97%, CA 96%. Small numbers of VME (0.9%), ME (1.4%) and mE (2.8%) were obtained. Overall, EA and CA rates for most drugs were above 90% for both genera. A few VME were found: a) teicoplanin and high-level streptomycin for E. faecalis, b) high-level gentamicin for E. faecium. The mean time to results (± SD) was 11.8 ± 0.9 h, with minor differences between streptococci and enterococci.
Conclusion
The Phoenix system emerged as an effective tool for quantitative AST. Panels based on dilution tests provided rapid and accurate MIC values with regard to clinically-relevant streptococcal and enterococcal species.
Background
Resistance of streptococci and enterococci to different classes of antimicrobials is increasing worldwide [1, 2]. Reduced susceptibility or resistance to penicillin and other β-lactams have been reported with particular frequency in the case of Streptococcus pneumoniae and viridans streptococci [1, 3]. Resistance to macrolides and fluoroquinolones has been described in viridans and β-hemolytic streptococci [4, 5]. Notably, glycopeptide resistance is an increasing threat among enterococcal species, especially Enterococcus faecium and E. faecalis [2].
Since clinical isolates of all the above species may produce severe infections (sepsis, meningitis, endocarditis), rapid and accurate identification (ID) and antimicrobial susceptibility testing (AST) are of great significance. For these bacteria, evaluation of drug susceptibility using disk diffusion methods can be problematic since the technique does not provide quantitative values. In this context, the Clinical Laboratory Standards Institute (CLSI; formerly NCCLS) established that S. pneumoniae isolates should not be reported as penicillin-resistant or intermediate based solely on an oxacillin zone ≤ 19 mm [6]. For these strains, minimum inhibitory concentration (MIC) values of penicillin and cefotaxime/ceftriaxone or meropenem are mandatory. Correct detection of vancomycin resistance among enterococci is also critical, since diffusion is difficult for large molecules and interpretative criteria are based on limited differences in the inhibition zone. MIC values may be of further assistance in discriminating enterococcal species carrying different van-type determinants.
Reproducibility of results, accuracy of species ID, precision of quantitative AST results, turnaround time, availability of data for epidemiological monitoring, and cost-effectiveness represent compelling reasons that support the choice of an automated system for clinical bacteriology. Systems of this type are now available for catalase-negative gram-positive cocci [7–9]. In the year 2000, Becton Dickinson (BD Diagnostic Systems, Sparks, MD) introduced the Phoenix Automated Microbiology System for ID and AST of gram-negative bacilli, staphylococci, and enterococci [10–13]. More recently, a panel for streptococcal species (SMIC/ID-2 panel) has been added [14, 15].
The present study was designed to evaluate the performance of the Phoenix system for the quantitative determination of drug susceptibility of streptococcal and enterococcal isolates carrying different resistance determinants. Time to results was also analyzed.
Results
Bacterial isolates are listed in Table 1. AST results are expressed as raw data (i.e., as results not interpreted by the expert system). For streptococcal isolates (Table 2), the overall essential agreement (EA; 91.9%) and categorical agreement (CA; 98.8%) between the Phoenix system and the Etest method were satisfactory. No VME were produced. Very low rates of major errors (ME; 0.1%) and minor errors (mE; 1.1%) were found. For enterococcal isolates (Table 3), the overall EA and CA were 97% and 96%, respectively. Small numbers of errors were found: VME (0.9%), ME (1.4%), mE (2.8%).
As shown in Tables 2 and 3, the overall performance for different antimicrobials was rather satisfactory (mean EA for streptococci ≥ 91%, for enterococci ≥ 97%). Poor performances were, however, noticed for streptococci in the case of clindamycin (EA, 67.3%) and erythromycin (EA, 88.6%), and for enterococci in the case of teicoplanin (EA, 87%). CA rates ≥ 97% and ≥ 94% were obtained for streptococci and enterococci, respectively. However, the low CA rate obtained in the case of linezolid for enterococci (89%) should be noticed.
Table 4 shows the AST results relative to different species or groups of streptococci. Mean CA values were as follows: S. pneumoniae 95%, β-hemolytic-streptococci 99.4%, and viridans streptococci 99.6%. With regard to beta-lactams, mE have been obtained only with S. pneumoniae: penicillin (12%), cefotaxime (2%), and cefepime (2%). Among S. pyogenes isolates, one ME for clindamycin was observed. As shown in Table 5, rather satisfactory CA values were obtained with E. faecalis (98.2%) and E. faecium (96.7%). Occasional VME have been produced for teicoplanin and HL-streptomycin in E. faecalis, and for HL-gentamicin in E. faecium. Due to the reduced numbers, results of isolates belonging to species other than E. faecalis and E. faecium (see Table 1) are not presented in Table 5. For these species, mean CA value was 88.9%; mE and ME were 7.2% and 5.3%, respectively. Errors were restricted to chloramphenicol, tetracycline, linezolid, and quinupristin-dalfopristin.
Time to results (TTR)
TTR values represent the overall time required for both ID and AST results. The mean TTR ± SD was 11.8 ± 0.9 h, with minor differences between streptococci (12.4 ± 2.2) and enterococci (11.4 ± 3.0). Among streptococci, β-hemolytic species showed the lowest TTR (S. agalactiae, 11.3 h; S. pyogenes, 11.6 h; S. dysgalactiae, 11.7 h). S. pneumoniae and viridans group streptococci were characterized by longer TTR (13.1 h and 13.3 h, respectively). Among enterococci, TTR was 9.9 ± 2.8 h for E. faecalis, and 11.7 ± 2.3 h for E. faecium.
Discussion
Over 70 streptococcal and enterococcal species have been implicated in animal and/or human infections [16, 17]. Of these, only a few cause important human infections (e.g., S. pyogenes, S. pneumoniae, S. agalactiae, E. faecalis, E. faecium, viridans streptococci).
Streptococcal and enterococcal isolates characterized by a wide spectrum of drug susceptibility have been investigated to assess the performance of the Phoenix system using both the new SMIC/ID-2 panel for streptococci and the established PMIC/ID-14 panel for enterococci. Both panels allowed testing clinically-relevant drugs: penicillins, third- and fourth-generation cephalosporins, macrolides, fluoroquinolones, glycopeptides, and the newer molecules linezolid and quinupristin-dalfopristin.
The results demonstrate the overall satisfactory performance of the Phoenix system. CA and EA values were consistently > 90% (98.8% and 91.9%, for streptococci; 96% and 97%, for enterococci). The lowest CA values were obtained with S. pneumoniae and with enterococci other than E. faecalis and E. faecium.
For streptococci, results in agreement with ours have been obtained by Japanese investigators [14, 15] that used SMIC/ID panels differing from those used by us. As compared to the automated Vitek system, the Phoenix appeared more accurate in determining MIC values for pneumococci [18]. Excellent results were obtained with both systems with regard to S. agalactiae [9]. Among enterococci, our results showed lower performances with respect to what has been reported by others using both the Phoenix and the Vitek systems [9, 12, 19]. In this respect, it should be noted that a high proportion (46/100) of our isolates carried glycopeptide resistance determinants and that 17/100 isolates were of species other than E. faecalis and E. faecium. Problems with AST results for the above species have also been reported by others using automated systems [20].
Unsatisfactory results have been obtained for clindamycin in the case of streptococci (EA, 67.3%): MIC values produced by the Phoenix were, in fact, usually lower than those obtained with the Etest. Investigated strains, however, showed MICs consistently within breakpoint values, thus limiting the clinical impact of the error. The reduced response to clindamycin was not related to the expression of resistance determinants (inducible/constitutive MLSB phenotype, or efflux; data not shown). Overall, small numbers of VME and ME were obtained. Most errors were classified as mE, due to minor differences in MIC values around breakpoint concentrations. Regarding ten enterococcal isolates, the Phoenix system gave MIC values of 4 μg/ml for linezolid, thus classifying the isolates as "intermediate". MIC values given by the Etest were instead of 2 μg/ml (i.e., "susceptible"). The finding appears to reflect a problem already noticed by Phoenix users with regard to this drug. Thus, increased linezolid MICs need to be verified using a reference method. Regarding penicillin, mE were encountered in 6/150 streptococci. In the latter cases, MICs were overestimated by the Phoenix system. Overall, the results of the present study were superior to those previously reported for both the Phoenix and the VITEK 2 system [14, 15, 18].
A mE rate < 10% is recommended for accepting AST results, and up to 3.0% ME and 1.5% VME are acceptable [21]. Overall, the results produced by the Phoenix were well within the recommended range. However, an ME rate of 5.3% was obtained for enterococci other than E. faecium and E. faecalis.
Conclusion
The Phoenix system appears an effective diagnostic tool for infections caused by streptococci and enterococci. Panels based on dilution tests were able to deliver adequate quantitative AST results in short times. This is especially relevant for severe infections caused by aggressive species such as S. pyogenes, S. pneumoniae, E. faecalis, and E. faecium. For these organisms, interpretation of susceptibility data varies according to the infection site and – in the case of large molecules that hardly diffuse in solid media (e.g., glycopeptides) – results obtained using dilution methods are to be preferred [20, 22].
Methods
Clinical isolates
Bacterial strains were obtained from routine clinical specimens at the Microbiology Laboratory of the Ospedale di Circolo, Varese (Italy). A total of 250 non-duplicate clinical isolates of gram-positive, catalase-negative cocci were studied. When needed, species ID was obtained by molecular methods [23]. The following species were investigated (Table 1): S. pneumoniae, Streptococcus pyogenes, Streptococcus agalactiae, Streptococcus dysgalactiae subsp. equisimilis, viridans group streptococci, E. faecium, E. faecalis, plus other streptococcal and enterococcal species. Test organisms included isolates characterized by different susceptibility levels to penicillin, macrolides, and fluoroquinolones. Among enterococci (n = 100), vancomycin resistance determinants were present in 46 isolates: vanA (33 isolates), vanB (2 isolates), vanC1 (7 isolates), and vanC2 (4 isolates). Fifty-four vancomycin-susceptible isolates were shown not to carry van-type determinants. Glycopeptide-resistance determinants (vanA, vanB, vanC-1, vanC-2, van-C-3, vanD, vanE, and vanG) were investigated by PCR following reported methods [24]. Isolates were stored at -70°C in Todd-Hewitt broth containing 20% glycerol. Before performing ID and AST assays, the isolates were passed twice on Mueller-Hinton agar containing 5% sheep blood (Oxoid SpA, Milan, Italy) to get them to an active growth stage following metabolic inactivity while frozen.
Phoenix AST
The Phoenix system uses two different panels for ID and AST of gram-positive cocci: the SMIC/ID-2 panel for streptococci, and the PMIC/ID-14 panel for enterococci and staphylococci. Panels have two separate sections. The left one contains ID substrates. The AST section (right side) contains three to eight concentrations of different drugs. Panel inoculation has been performed according to the manufacturer's instructions. As recommended, the same ID broth has been used for both panel types. Two different broth types are instead needed for AST. The AST-S broth containing growth supplements plus one drop of the Phoenix AST-S indicator is intended for streptococci. The AST broth plus one drop of the Phoenix AST indicator is intended for enterococci, staphylococci, and gram-negative bacteria.
Bacterial isolates have been obtained from fresh overnight cultures, suspended in the ID broth and adjusted to the 0.5 McFarland standard using a dedicated nephelometer (CrystalSpec, BD). Twenty-five microliters of the ID suspension have been used to inoculate AST broths, resulting in the final inoculum density of approximately 5 × 105 CFU/ml. Inoculated AST broths have been poured into the appropriate sections of each panel. Panels have been sealed, logged, and loaded into the instrument at 35°C. Kinetic, colorimetric and fluorescent signals have been collected every 20 min until ID and AST results were completed.
Comparator AST method
MICs have been determined by the Etest method (AB Biodisk, Solna, Sweden) according to manufacturer's instructions (preparation of inoculum, plating, strip application, reading of results). Streptococcal isolates have been cultured on Mueller-Hinton agar plates supplemented with 5% sheep blood (Oxoid). Enterococcal isolates have been tested on Mueller-Hinton agar plates (BD). Vancomycin has been tested on brain heart infusion plates (Oxoid). MIC values have been interpreted according to CLSI criteria [6]. Different drug panels have been tested for streptococci and enterococci: a) streptococci (penicillin, erythromycin, clindamycin, cefotaxime, cefepime, levofloxacin, vancomycin, chloramphenicol); b) enterococci [ampicillin, vancomycin, teicoplanin, high-level (HL) streptomycin, HL gentamicin, chloramphenicol, quinupristin-dalfopristin, tetracycline, linezolid].
Analysis of data and discrepancies
AST results are reported as raw data, i.e. without interpretation by the expert system. Results have been categorized as susceptible (S), intermediate (I), or resistant (R) according to CLSI [6]. MIC values produced by the Phoenix method (TEST) have been compared to those obtained with the Etest method (COMP). The following definitions have been adopted: 1) EA (i.e., MIC values of TEST panel equal to or within ± 1 dilution of the COMP value); 2) CA (i.e., TEST and COMP MIC values agree using insterpretative CLSI criteria); 3) mE (i.e., COMP is S or R and TEST is I; alternatively, COMP is I and TEST is S or R); 4) ME (i.e., COMP is S and TEST is R; the percentage of major errors has been calculated only for susceptible isolates); 5) VME (i.e., COMP is R and TEST is S; the percentage of very major errors has been calculated only for resistant isolates). In the case of ME or VME, the organism has been re-tested with both the Phoenix and the Etest. When differences persisted, the results have been considered discordant.
Quality controls
As controls, the following strains have been included in each run: S. pneumoniae ATCC 49619, S. agalactiae ATCC 13813, and E. faecalis ATCC 29212.
Abbreviations
- (AST):
-
Antimicrobial susceptibility testing
- (CA):
-
Categorical agreement
- (CLSI):
-
Clinical Laboratory Standards Institute
- (COMP):
-
Comparator method
- (EA):
-
Essential agreement
- (ID):
-
Identification
- (I):
-
Intermediate
- (ME):
-
Major errors
- MIC):
-
Minimal inhibitory concentration
- (mE):
-
Minor errors
- (R):
-
Resistant
- (S):
-
Susceptible
- (TEST):
-
Test method
- (TTR):
-
Time to results
- (VME):
-
Very major errors
References
Reinert RR, Reinert S, van der Linden M, Cil MY, Al-Lahham A, Appelbaum P: Antimicrobial susceptibility of Streptococcus pneumoniae in eight European countries from 2001 to 2003. Antimicrob Agents Chemother. 2005, 49: 2903-2913. 10.1128/AAC.49.7.2903-2913.2005.
Treitman AN, Yarnold PR, Warren J, Noskin GA: Emerging incidence of Enterococcus faecium among hospital isolates (1993 to 2002). J Clin Microbiol. 2005, 43: 462-463. 10.1128/JCM.43.1.462-463.2005.
Prabhu RM, Piper KE, Baddour LM, Steckelberg JM, Wilson WR, Patel R: Antimicrobial susceptibility patterns among viridans group streptococcal isolates from infective endocarditis patients from 1971 to 1986 and 1994 to 2002. Antimicrob Agents Chemother. 2004, 48: 4463-4465. 10.1128/AAC.48.11.4463-4465.2004.
Gordon KA, Beach ML, Biedenbach DJ, Jones RN, Rhomberg PR, Mutnick AH: Antimicrobial susceptibility patterns of β-hemolytic and viridans group streptococci: report from the SENTRY Antimicrobial Surveillance Program (1997–2000). Diagn Microbiol Infect Dis. 2002, 43: 157-162. 10.1016/S0732-8893(02)00374-7.
Biedenbach DJ, Toleman MA, Walsh TR, Jones RN: Characterization of fluoroquinolone-resistant β-hemolytic Streptococcus spp. Isolated in North America and Europe including ffthe first report of fluoroquinolone-resistant Streptococcus dysgalactiae subspecies equisimilis: Report from the SENTRY Antimicrobial Surveillance Program (1997–2004). Diagn Microbiol Infect Dis. 2006, 55: 119-127. 10.1016/j.diagmicrobio.2005.12.006.
Clinical and Laboratory Standards Institute: Performance standards for antimicrobial susceptibility testing. Sixteenth informational supplement M100-S16. Clinical and Laboratory Standards Institute, Wayne, PA. 2006
Eisner A, Gorkiewicz G, Feierl G, Leitner E, Köfer J, Kessler HH, Marth E: Identification of glycopeptide-resistant enterococci by VITEK 2 system and conventional and real-time polymerase chain reaction. Diagn Microbiol Infect Dis. 2005, 53: 17-21. 10.1016/j.diagmicrobio.2005.04.001.
Gavin PJ, Warren JR, Obias AA, Collins SM, Peterson LR: Evaluation of the Vitek 2 system for rapid identification of clinical isolates of gram-negative bacilli and members of the family Streptococcaceae. Eur J Clin Microbiol Infect Dis. 2002, 21: 869-874.
Ligozzi M, Bernini C, Bonora MG, de Fatima M, Zuliani J, Fontana R: Evaluation of the VITEK 2 system for identification and antimicrobial susceptibility testing of medically relevant gram-positive cocci. J Clin Microbiol. 2002, 40: 1681-1686. 10.1128/JCM.40.5.1681-1686.2002.
Donay JL, Mathieu D, Fernandes P, Prégermain C, Bruel P, Wargnier A, Casin I, Weill FX, Lagrange PH, Herrmann JL: Evaluation of the automated Phoenix system for potential routine use in the clinical microbiology laboratory. J Clin Microbiol. 2004, 42: 1542-1546. 10.1128/JCM.42.4.1542-1546.2004.
Endimiani A, Luzzaro F, Tamborini A, Lombardi G, Elia V, Belloni R, Toniolo A: Identification and antimicrobial susceptibility testing of clinical isolates of nonfermenting gram-negative bacteria by the Phoenix automated microbiology system. New Microbiol. 2002, 25: 323-329.
Fahr AM, Eigner U, Armbrust M, Caganic A, Dettori G, Chezzi C, Bertoncini L, Benecchi M, Menozzi MG: Two-center collaborative evaluation of the performance of the BD Phoenix automated microbiology system for identification and antimicrobial susceptibility testing of Enterococcus spp. and Staphylococcus spp. J Clin Microbiol. 2003, 41: 1135-1142. 10.1128/JCM.41.3.1135-1142.2003.
O'Hara CM: Manual and automated instrumentation for identification of Enterobacteriaceae and other aerobic gram-negative bacilli. Clin Microbiol Rev. 2005, 18: 147-162. 10.1128/CMR.18.1.147-162.2005.
Hirakata Y, Matsuda J, Nakano M, Hayashi T, Tozaka S, Takezawa T, Takahashi H, Higashiyama Y, Miyazaki Y, Kamihira S, Kohno S: Evaluation of the BD Phoenix automated microbiology system SMIC/ID panel for identification and antimicrobial susceptibility testing of Streptococcus spp. Diagn Microbiol Infect Dis. 2005, 53: 169-173. 10.1016/j.diagmicrobio.2005.06.003.
Kanemitsu K, Kunishima H, Inden K, Hatta M, Harigae H, Ishizawa K, Kaku M: Evaluation of the BD Phoenix SMIC/ID, a new streptococci identification and antimicrobial susceptibility panel, for potential routine use in a university-based clinical microbiology laboratory. Diagn Microbiol Infect Dis. 2005, 53: 101-105. 10.1016/j.diagmicrobio.2005.05.016.
Facklam R: What happened to the streptococci: overview of taxonomic and nomenclature changes. Clin Microbiol Rev. 2002, 15: 613-630. 10.1128/CMR.15.4.613-630.2002.
Teixeira LM, Facklam RR:Enterococcus. Manual of Clinical Microbiology. Edited by: Murray PR, Barron EJ, Jorgensen JH, Pfaller MA, Yolken RH. 2003, American Society for Microbiology, Washington, DC, 422-433. 8,
Goessens WHF, Lemmens-den Toom N, Hageman J, Hermans PWM, Sluijter M, de Groot R, Verbrugh HA: Evaluation of the Vitek 2 system for susceptibility testing of Streptococcus pneumoniae isolates. Eur J Clin Microbiol Infect Dis. 2000, 19: 618-622. 10.1007/s100960000332.
Eigner U, Schmid A, Wild U, Bertsch D, Fahr A-M: Analysis of the comparative workflow and performance characteristics of the VITEK 2 and Phoenix systems. J Clin Microbiol. 2005, 43: 3829-3834. 10.1128/JCM.43.8.3829-3834.2005.
Abele-Horn M, Hommers L, Trabold R, Frosch M: Validation of VITEK 2 4.01 software for detection, identification, and classification of glycopeptide-resistant enterococci. J Clin Microbiol. 2006, 44: 71-76. 10.1128/JCM.44.1.71-76.2006.
National Committee for Clinical Laboratory Standards: Development of in vitro susceptibility testing criteria and quality control parameters. Approved guideline-second edition. Document M23-A2, National Committee for Clinical Laboratory Standards, Wayne, Pa. 2001
Kohner PC, Patel R, Uhl JR, Garin KM, Hopkins MK, Wegener LT, Cockerill FR: Comparison of agar dilution, broth microdilution, E-test, disk diffusion, and automated Vitek methods for testing susceptibilities of Enterococcus spp. to vancomycin. J Clin Microbiol. 1997, 35: 3258-3263.
Brigante G, Luzzaro F, Bettaccini A, Lombardi G, Meacci F, Pini B, Stefani S, Toniolo A: Use of the Phoenix automated system for identification of Streptococcus and Enterococcus spp. J Clin Microbiol. 2006, 44: 3263-3267. 10.1128/JCM.00299-06.
Depardieu F, Perichon B, Courvalin P: Detection of the van alphabet and identification of enterococci and staphylococci at the species level by multiplex PCR. J Clin Microbiol. 2004, 42: 5857-5860. 10.1128/JCM.42.12.5857-5860.2004.
Acknowledgements
The excellent technical contribution of Vito Elia, Mirta Broggi, Clara De Bortoli, Pasquale Abbate, Riccardo Fusar-Imperatore, Sergio Gallazzi, Paola Caputo, Rosalia Bontà, Francesco Tucci and Nunzia Vocino is gratefully acknowledged. This work was supported by grants from the Italian Ministry Education, University and Scientific Research (MIUR, Rome), the Italian Ministry of Health (IMH, Rome), and Becton Dickinson (Milan, Italy).
Author information
Authors and Affiliations
Corresponding author
Additional information
Authors' contributions
GRB and BP carried out drug susceptibility tests and species identification; FAL and GL provided bacterial isolates together with clinical data; GS carried out molecular studies; AQT conceived and coordinated the study. All authors read and approved the final manuscript.
Rights and permissions
Open Access This article is published under license to BioMed Central Ltd. This is an Open Access article is distributed under the terms of the Creative Commons Attribution License ( https://creativecommons.org/licenses/by/2.0 ), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
About this article
Cite this article
Brigante, G.R., Luzzaro, F.A., Pini, B. et al. Drug susceptibility testing of clinical isolates of streptococci and enterococci by the Phoenix automated microbiology system. BMC Microbiol 7, 46 (2007). https://doi.org/10.1186/1471-2180-7-46
Received:
Accepted:
Published:
DOI: https://doi.org/10.1186/1471-2180-7-46