Skip to main content

Table 2 Comparative evaluation of DENV serotype-specific RT-LAMP assay with RT-PCR, and real-time PCR for detection of DENV stains and the patients’ serum samples

From: Development of reverse-transcription loop-mediated isothermal amplification assay for rapid detection and differentiation of dengue virus serotypes 1–4

Type of case

Virus serotype

No. of samples Virus isolation

No. of samples positive by:

RT-LAMP

RT-PCR

Real-time PCR

Virus stains

DEN-1

20

20

18

20

 

DEN-2

30

30

27

30

 

DEN-3

15

15

15

15

 

DEN-4

15

15

14

15

Total

 

80

80

74

80

Accuracy (%)

  

100

92.5

100

Patients’ serum samples

DEN-1

50

50

41

44

 

DEN-2

60

59

52

54

 

DEN-3

40

40

35

38

 

DEN-4

40

39

33

36

Total

 

190

188

161

172

Accuracy (%)

  

98.95

84.74

90.53

Healthy

 

20

0

0

0

Total

 

20

0

0

0

False positive rate (%)

  

0

0

0