Skip to main content

Table 2 The intermediate precision and accuracy of the developed RT-qPCR infectivity assay is determined

From: Evaluation of infectious titer in a candidate HSV type 2 vaccine by a quantitative molecular approach

Assay #

RT-qPCR (pfu)

RT-qPCR (log_pfu)

Plaque assay

Accuracy%

CV%

(Mean from 30 assays)

1

1.50E + 07

16.52

1.41E + 07

106.38

 

2

1.63E + 07

16.60

115.60

3

1.45E + 07

16.48

102.34

4

1.70E + 07

16.64

120.57

5

1.54E + 07

16.54

109.22

6

1.31E + 07

16.38

 

92.91

9.19

  1. RT-qPCR infectivity assay was performed six times by two analysts on different days. The accuracy of the assay was calculated by evaluating the percentages of values obtained by RT-qPCR infectivity assay versus the expected infectious titre values by plaque assay. The CV% from the six independent assays is also determined.