Skip to main content

Table 2 Relative deviation of zone diameter values and resulting discrepancies of the Sirscan (on-screen adjusted) and manual calliper measurements

From: Standardisation of disk diffusion results for antibiotic susceptibility testing using the sirscan automated zone reader

 

Relative deviation of zone diameters values

Discrepancies

(% of all measurements)

(% of all Sirscan measurements)

 

Sirscan < calliper

Sirscan = calliper

Sirscan > calliper

minor

major

very major

Gram-negative rods

19

45

36

1.27

0

0

Staphylococcus spp.

27

37

36

0.94

0

0

Enterococcus spp.

53

35

12

0

0

0

  1. For discrepancy analysis manual calliper measurements were regarded as the gold standard. Sirscan values were on-screen adjusted by an experienced person as recommended by the manufacturer.